The authors describe a new technique that uses simultaneous transmission and scatter SPECT imaging for determination of segmented attenuation /spl mu/-maps. These /spl mu/-maps could provide the basis for attenuation correction and quantification of radiopharmaceutical uptake in cardiac SPECT. For transmission imaging, a line source and a fan-beam collimator on one head of a triple-head SPECT system are used. For scatter imaging, two external ring sources and parallel collimators on one or two of the other heads are used. The fan-beam transmission data are truncated, while the parallel-beam data are not. The proposed technique uses both sets of tomographic data to complement each other and to obtain a composite image. A /spl mu/-map with three segmented regions of distinct /spl mu/-values could be derived from the composite image. This final /spl mu/-map can then be used as the basis for attenuation correction.>
No takes yet. Share an insight, caveat, or question.
Lončarić et al. (1994) studied this question.