Answers are grounded in this paper and cite their sources.
PURPOSE: To determine if LASIK using a 130-microm microkeratome head (thin-flap LASIK) is as visually effective and safe as when using a 160-microm head. METHODS: A study was performed comparing postoperative day 1 uncorrected visual acuity (UCVA) and flap complications in eyes undergoing myopic LASIK with a 130-microm head versus a 160-microm head using the BD K-3000 microkeratome (BD Ophthalmic Systems, Waltham, Mass). RESULTS: The mean preoperative myopia in the 155 eyes of 80 patients in the 130-microm head group was -5.00+/-2.53 diopters (D) compared to -3.78+/-1.73 D in the 279 eyes of 148 patients in the 160-microm head group. The groups were otherwise similar in terms of age, preoperative cylinder, and best spectacle-corrected visual acuity. The geometric mean postoperative day 1 UCVA was 20/25 in the 130-microm head group compared to 20/26 in the 160-microm head group. The only flap complication in the series was a single partial flap in the 160-microm head group. CONCLUSIONS: LASIK with a 130-microm head (thin-flap LASIK) is as visually effective and safe as when using a 160-microm head with the BD K-3000 microkeratome.
No takes yet. Share an insight, caveat, or question.
Paul J. Dougherty (2005) studied this question.