Key result
A proposed cascaded 4-stage adaptive noise canceller employing LMMN, LMF, LMMN, and LMF algorithms achieved an SNR improvement of 12.7319 dBs for removing multiple artifacts from ECG signals.
Why the study?
Previous cascaded multistage adaptive noise cancellers for ECG artifact removal used the same adaptive algorithm across all stages, rather than algorithms tailored to specific artifacts.
A novel cascaded 4-stage adaptive noise canceller using a combination of specific algorithms effectively removes multiple artifacts from ECG signals, improving SNR by 12.73 dB.
May aid ECG artifact removal in monitoring; leaves open clinical validation before practice adoption.
Although cascaded multistage adaptive noise cancellers have been employed before by researchers for multiple artifact removal from the ElectroCardioGram (ECG) signal, they all used the same adaptive algorithm in all the cascaded multi-stages for adjusting the adaptive filter weights. In this paper, we propose a cascaded 4-stage adaptive noise canceller for the removal of four artifacts present in the ECG signal, viz. baseline wander, motion artifacts, muscle artifacts, and 60 Hz Power Line Interference (PLI). We have investigated the performance of eight adaptive algorithms, viz. Least Mean Square (LMS), Least Mean Fourth (LMF), Least Mean Mixed-Norm (LMMN), Sign Regressor Least Mean Square (SRLMS), Sign Error Least Mean Square (SELMS), Sign-Sign Least Mean Square (SSLMS), Sign Regressor Least Mean Fourth (SRLMF), and Sign Regressor Least Mean Mixed-Norm (SRLMMN) in terms of Signal-to-Noise Ratio (SNR) improvement for removing the aforementioned four artifacts from the ECG signal. We employed the LMMN, LMF, LMMN, LMF algorithms in the proposed cascaded 4-stage adaptive noise canceller to remove the respective ECG artifacts as mentioned above. We succeeded in achieving an SNR improvement of 12.7319 dBs. The proposed cascaded 4-stage adaptive noise canceller employing the LMMN, LMF, LMMN, LMF algorithms outperforms those that employ the same algorithm in the four stages. One unique and powerful feature of our proposed cascaded 4-stage adaptive noise canceller is that it employs only those adaptive algorithms in the four stages, which are shown to be effective in removing the respective ECG artifacts as mentioned above. Such a scheme has not been investigated before in the literature.
No takes yet. Share an insight, caveat, or question.
Faiz et al. (2022) studied ECG artifacts. Cascaded 4-stage adaptive noise canceller employing LMMN, LMF, LMMN, LMF algorithms vs. Cascaded multi-stages using the same adaptive algorithm was evaluated on Signal-to-Noise Ratio (SNR) improvement. A proposed cascaded 4-stage adaptive noise canceller employing LMMN, LMF, LMMN, and LMF algorithms achieved an SNR improvement of 12.7319 dBs for removing multiple artifacts from ECG signals.