A Characteristic feature of the radiation field of a radium applicator or implant is the high gradient of dosage perpendicular to the radium bearing surface. Parallel to that surface the dosage level is uniform within certain limits depending upon the arrangement of the radium sources. Any physical measurement of the γ-ray output of such a radium distribution therefore necessitates the use of a very small ionisation chamber, if the recorded ionisation is to give a measure of the dosage rate at the geometrical centre of the chamber. If the dimensions of the ionisation chamber are of the same order as the radium source distances, inverse square law errors may be expected to arise owing to the non-uniformity of the radiation intensity across the chamber. Condenser ionisation chambers commonly used for γ-ray measurements have dimensions of about one centimetre, or somewhat less, and in many dose measurements on moulds and implants radium source distances of this magnitude are involved. Some investigation has therefore been made of the effect of a finite condenser size on measurements made in close proximity to radium sources and of the limitations thus imposed on the use of standard types of condenser chambers. This question has been referred to briefly in a recent paper1 and some of the results of the calculations which follow were then stated. Some experimental work published by Mayneord and Roberts2 bears upon this question of condenser size and their results agree well with the present calculations.