Answers are grounded in this paper and cite their sources.
A Johann-type reflection spectrometer with a wavelength resolution of 15000 at 0.229 nm (CrK alpha ) is described. The diffracting crystal is a (310)-quartz plate with 50*30 mm 2 total and useful area, bent to a radius of 1380 mm. The use of a novel, simple crystal bending device and of a position-sensitive proportional counter makes this instrument interesting also for other domains of X-ray spectroscopy. Results of extended calculations on diffraction properties of quartz as a function of diffraction plane index and wavelength are also presented.
No takes yet. Share an insight, caveat, or question.
Platz et al. (1981) studied this question.