Key result
A standardized 12-lead fetal electrocardiogram was successfully recorded in 98.3% of healthy mid-pregnancy fetuses, revealing a right inferiorly oriented median electrical heart axis.
Why the study?
Mid-pregnancy fetal heart ultrasound is highly dependent on operator skill, fetal position, and maternal BMI, making an additional, less experience-dependent tool such as fetal ECG desirable.
Is it feasible to record a standardized 12-lead fetal ECG in mid-pregnancy to assess the fetal heart and detect congenital heart disease?
Cross-Sectional (n=328)
Yes
Is it feasible to record a standardized 12-lead fetal ECG in mid-pregnancy to assess the fetal heart and detect congenital heart disease?
Non-invasive standardized 12-lead fetal electrocardiography is highly feasible in mid-pregnancy and may serve as a valuable adjunct to ultrasound for detecting congenital heart disease.
Answers are grounded in this paper and cite their sources.
High feasibility of 12-lead fetal ECG in mid-pregnancy supports further study; leaves open its adjunctive role for congenital heart disease detection.
INTRODUCTION: The examination of the fetal heart in mid-pregnancy is by ultrasound examination. The quality of the examination is highly dependent on the skill of the sonographer, fetal position and maternal body mass index. An additional tool that is less dependent on human experience and interpretation is desirable. The fetal electrocardiogram (ECG) could fulfill this purpose. We aimed to show the feasibility of recording a standardized fetal ECG in mid-pregnancy and explored its possibility to detect congenital heart disease (CHD). MATERIALS AND METHODS: Women older than 18 years of age with an uneventful pregnancy, carrying a healthy singleton fetus with a gestational age between 18 and 24 weeks were included. A fetal ECG was performed via electrodes on the maternal abdomen. After removal of interferences, a vectorcardiogram was constructed. Based on the ultrasound assessment of the fetal orientation, the vectorcardiogram was rotated to standardize for fetal orientation and converted into a 12-lead ECG. Median ECG waveforms for each lead were calculated. RESULTS: 328 fetal ECGs were recorded. 281 were available for analysis. The calculated median ECG waveform showed the electrical heart axis oriented to the right and inferiorly i.e. a negative QRS deflection in lead I and a positive deflection in lead aVF. The two CHD cases show ECG abnormalities when compared to the mean ECG of the healthy cohort. DISCUSSION: We have presented a method for estimating a standardized 12-lead fetal ECG. In mid-pregnancy, the median electrical heart axis is right inferiorly oriented in healthy fetuses. Future research should focus on fetuses with congenital heart disease.
No takes yet. Share an insight, caveat, or question.
Lempersz et al. (2020) conducted a cross-sectional in Healthy fetus in mid-pregnancy (n=328). Standardized 12-lead fetal electrocardiogram was evaluated on Feasibility of recording a standardized fetal ECG and determining median ECG waveform. A standardized 12-lead fetal electrocardiogram was successfully recorded in 98.3% of healthy mid-pregnancy fetuses, revealing a right inferiorly oriented median electrical heart axis.