Key result
Managed VP, search AV, and fixed long AV pacing strategies produced similar mean atrial and ventricular pacing percentages with no statistically significant differences between groups.
Why the study?
Do different pacing algorithms (managed VP, search AV, fixed long AV) differ in reducing unnecessary ventricular pacing in patients with sick sinus syndrome?
Observational (n=56)
No
Do different pacing algorithms (managed VP, search AV, fixed long AV) differ in reducing unnecessary ventricular pacing in patients with sick sinus syndrome?
Different pacing algorithms (managed VP, search AV, and fixed long AV) are equally effective in reducing unnecessary ventricular pacing in patients with sick sinus syndrome.
Answers are grounded in this paper and cite their sources.
May support algorithm interchangeability in sick sinus syndrome; leaves open need for randomized confirmation.
AIMS: Reduction of unnecessary ventricular pacing (uVP) is an essential component in the treatment strategy in any pacing population in general. The aim of this study was to evaluate the efficacy of different algorithms to reduce uVP in an adult population with sick sinus syndrome (SSS) treated outside of clinical trials. Evaluation of the relationship between different types of pacing algorithms and clinical outcomes is also provided. METHODS AND RESULTS: This was a single-centre, observational, parallel study, based on retrospective analysis of the Arrhythmology Cardiology Center of Armenia electronic clinical database. This study evaluated atrial pacing percentage (AP%), ventricular pacing percentage (VP%), and the incidence of atrial high rate episodes in 56 patients with SSS using three different pacing strategies: managed VP, search atrioventricular (AV), and fixed long AV. We did not find statistically significant differences in the amount of VP between the groups. Although the atrial high rate percentage (AHR%) tended to be higher in the fixed long AV group, this difference was not statistically significant. Mean VP% and AP% were similar in all three groups. CONCLUSIONS: In our study, all three programmed strategies produced the same mean AP% and VP%, and were equally efficient in uVP reduction. There was no relationship between chosen algorithms and the incidence of pacemaker syndrome, hospitalizations, or change in New York Heart Association class. The percentage of AHR was not associated with pacing strategy or co-morbidities but showed borderline correlation with left atrial size.
No takes yet. Share an insight, caveat, or question.
Poghosyan et al. (2012) conducted an observational in Sick sinus syndrome (n=56). Pacing strategies (managed VP, search AV, and fixed long AV) vs. Compared against each other was evaluated on Atrial pacing percentage, ventricular pacing percentage, and incidence of atrial high rate episodes. Managed VP, search AV, and fixed long AV pacing strategies produced similar mean atrial and ventricular pacing percentages with no statistically significant differences between groups.