A versatile orthogonal-coil radio frequency (RF) probe suitable for detecting the resonant frequency of miniature implantable passive sensors has been designed and tested. The probe sensitivity has been tested using printed-circuit spiral inductors of various sizes (3-15 mm) in series with discrete surface-mount capacitors designed to resonate over a range of frequencies (50-200 MHz). Close agreement between theoretical calculations and experimental results has been obtained. An equation is derived for transmit/receive (T/R) isolation that agrees with experimental measurements over the frequency range 1-500 MHz. The probe includes an additional coil to compensate for the effect of eddy currents in the human body on the probe. T/R isolation of at least 90 dB over the frequency range 1-100 MHz can be achieved when the probe is placed in close proximity to the human body.
No takes yet. Share an insight, caveat, or question.
Talman et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: