Key result
The Left Ventricular Capture Management algorithm was clinically equivalent to manual testing, with 99.7% of automatic tests within one programming step of manual tests.
Why the study?
Does the LVCM algorithm accurately measure LV stimulation thresholds compared to manual testing in patients with CRT-D systems?
Does the LVCM algorithm accurately measure LV stimulation thresholds compared to manual testing in patients with CRT-D systems?
The LVCM algorithm provides safe, accurate, and reliable automatic measurement of LV stimulation thresholds, equivalent to manual testing, which may improve device longevity and LV capture maintenance.
Answers are grounded in this paper and cite their sources.
Supports automated LV threshold testing in CRT-D; leaves open randomized confirmation of longevity benefits.
BACKGROUND: The stimulation thresholds of left ventricular (LV) leads tend to be less reliable than conventional leads. Cardiac resynchronization therapy (CRT) requires continuous capture of both ventricles. OBJECTIVE: The purpose of this study is to evaluate a novel algorithm for the automatic measurement of the stimulation threshold of LV leads in cardiac resynchronization systems. METHODS: We enrolled 134 patients from 18 centers who had existing CRT-D systems. Software capable of automatically executing LV threshold measurements was downloaded into the random access memory (RAM) of the device. The threshold was measured by pacing in the left ventricle and analyzing the interventricular conduction sensed in the right ventricle. Automatic LV threshold measurements were collected and compared with manual LV threshold tests at each follow-up visit and using a Holter monitor system that recorded both the surface electrocardiograph (ECG) and continuous telemetry from the device. RESULTS: The proportion of Left Ventricular Capture Management (LVCM) in-office threshold tests within one programming step of the manual threshold test was 99.7% (306/307) with a two-sided 95% confidence interval of (98.2%, 100.0%). The algorithm measured the threshold successfully in 96% and 97% of patients after 1 and 3 months respectively. Holter monitor analysis in a subset of patients revealed accurate performance of the algorithm. CONCLUSION: This study demonstrated that the LVCM algorithm is safe, accurate, and highly reliable. LVCM worked with different types of leads and different lead locations. LVCM was demonstrated to be clinically equivalent to the manual LV threshold test. LVCM offers automatic measurement, output adaptation, and trends of the LV threshold and should result in improved ability to maintain LV capture without sacrificing device longevity.
No takes yet. Share an insight, caveat, or question.
Crossley et al. (2007) studied Existing CRT-D systems (n=134). Left Ventricular Capture Management (LVCM) algorithm vs. Manual LV threshold test was evaluated on Proportion of LVCM in-office threshold tests within one programming step of the manual threshold test (95% CI 98.2-100.0). The Left Ventricular Capture Management algorithm was clinically equivalent to manual testing, with 99.7% of automatic tests within one programming step of manual tests.