Key result
Determining pacing threshold with increasing output yielded an average threshold 0.20 V/0.20 mA higher than with decreasing output, regardless of pacemaker indication or electrode status.
Why the study?
Does pacing threshold determination with increasing output yield different values compared to decreasing output in patients with pacemakers?
Observational (n=44)
Does pacing threshold determination with increasing output yield different values compared to decreasing output in patients with pacemakers?
Pacing thresholds measured with increasing output are consistently higher than those measured with decreasing output, regardless of pacemaker indication or electrode age.
Answers are grounded in this paper and cite their sources.
Threshold measurement order may bias values in device follow-up; leaves open standardization effects on programming and longevity.
The study was made in 44 patients, of which 23 received a new endocardial electrode and 21 were investigated during pulse-generator replacement. Indication for pacemaker therapy was sino-atrial bradycardia in 12 patients and atrio-ventricular block in 32 patients. When the pacing threshold was determined with decreasing output it was found to be 0.6V/0.7 mA for new implantations and 2.0 V/2.0 mA with pulse generator replacement. When determined with increasing output the threshold was found to be an average 0.20 V/0.20 mA higher than when determined with decreasing output (threshold interval). The maximal difference observed was 0.70 V/0.55 mA. This threshold interval was of the same size irrespective of pacemaker indication and spontaneous activity and whether the electrode was newly implanted or not. Pacemaker ventricular block of Mobitz type I with Wenckebach periodicity was never observed.
No takes yet. Share an insight, caveat, or question.
Sylvén et al. (1982) conducted an observational in Sino-atrial bradycardia or atrio-ventricular block (n=44). Increasing output for pacing threshold determination vs. Decreasing output was evaluated on Pacing threshold (average 0.20 V/0.20 mA higher). Determining pacing threshold with increasing output yielded an average threshold 0.20 V/0.20 mA higher than with decreasing output, regardless of pacemaker indication or electrode status.