Key result
The AutoCapture algorithm in a VVI pacemaker can produce two pacing spikes on the QRS complex due to a high output back-up pulse after loss of capture, simulating a sensing failure.
Case Report (n=1)
Understanding the AutoCapture algorithm is essential to avoid misdiagnosing pacemaker malfunction when two pacing spikes are seen on a QRS complex.
Answers are grounded in this paper and cite their sources.
May avert misdiagnosis of pacemaker malfunction; extends device algorithm troubleshooting knowledge.
AutoCapture algorithm helps the pacemakers to automatically adjust output close to the pacing threshold. This algorithm monitors the evoked response signal of myocardial depolarization, and delivers a high output back-up pulse if there is a loss of capture. This case, with two pacemaker spikes on the QRS complexes in a patient with VVI pacemaker, simulating a sensing failure, demands clear understanding of AutoCapture function before labeling the pacemaker malfunction.
No takes yet. Share an insight, caveat, or question.
Choudhary et al. (2012) conducted a case report in VVI pacemaker with two pacing spikes on QRS complex (n=1). AutoCapture algorithm was evaluated. The AutoCapture algorithm in a VVI pacemaker can produce two pacing spikes on the QRS complex due to a high output back-up pulse after loss of capture, simulating a sensing failure.