Key result
In pacemaker-dependent subjects, an optimized sensor setting did not increase maximal workload (133 W vs 129 W) or augment cardiac output compared to a default setting.
Why the study?
Does an optimized pacemaker sensor setting improve cardiac output and cerebral blood velocity during exercise in pacemaker-dependent subjects?
Population
Pacemaker-dependent subjects with complete heart block and preserved myocardial function
Comparison
Pacemaker set to 'optimized' sensor setting… vs Pacemaker set to 'default' sensor setting during…
Authors
Loading...
No benefit on workload or cardiac output with optimized settings; leaves open effects on cerebral perfusion in pacemaker-dependent patients.
Does an optimized pacemaker sensor setting improve cardiac output and cerebral blood velocity during exercise in pacemaker-dependent subjects?
Absolute Event Rate: 133% vs 129%
In pacemaker-dependent patients with complete heart block, a more aggressive rate-responsive pacing setting during exercise does not improve cardiac output or cerebral perfusion compared to default settings.
Bogert et al. (2005) studied Pacemaker-dependent with complete heart block. Optimized sensor setting (OSS) vs. Default sensor setting (DSS) was evaluated on Maximal workload. In pacemaker-dependent subjects, an optimized sensor setting did not increase maximal workload (133 W vs 129 W) or augment cardiac output compared to a default setting.