Answers are grounded in this paper and cite their sources.
The transport efficiency of the particle beam liquid chromatography/mass spectrometer interface is influenced by analyte concentration contributing to a widely reported non-linearity. In this work, coeluting, isotope-labeled internal standards were investigated as 'carriers' to improve the transport efficiency and linearity. Three styrene metabolites--mandelic, phenylglyoxylic and hippuric acids--and their pentadeutero analogs were separated by reversed-phase liquid chromatography (LC) with an ammonium acetate-acetonitrile mobile phase. Selected positive ions produced by electron ionization were monitored to generate particle beam LC/MS calibration curves. The present study demonstrates that particle beam LC/MS not only is non-linear, but also is subject to a matrix effect presumably by the same mechanism responsible for non-linearity. Coeluting, isotope-labeled internal standards were ineffective at linearizing the particle beam liquid chromatograph/mass spectrometer detector response. Isotope dilution quantification, however, compensates for variable transport efficiencies, linearizes calibration and compensates for the matrix effect, affording reliable quantification of the styrene metabolites.
No takes yet. Share an insight, caveat, or question.
Brown et al. (1991) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: