We describe an automated assay for progesterone (P4) in human serum and plasma with the Abbott AxSYM random-access immunoassay analyzer. In this one-step competitive assay, P4 immobilized onto latex microparticles competes with sample P4 for binding to a conjugate of alkaline phosphatase (AP) and anti-P4 antibody. Total CVs ranged from 3.4% to 8.2% in multiple precision studies conducted according to the 20-day NCCLS EP5-T protocol. The detection limit (zero calibrator + 2 SD) was 0.10 microg/L across 36 experiments. Values for diluted samples were 83-116% of expected. Recovery of P4 added to serum specimens was 92-115%. Cross-reactivities with 43 natural and synthetic steroids were 0-6.3%. No significant interference was detected from bilirubin, protein, erythrocytes, hemoglobin, triglycerides, or cholesterol. In a multisite correlation study, AxSYM P4 results compared well with results from a commercial RIA method (n = 1156; r = 0.976; slope = 1.03; y-intercept = 0.04). Assay throughput is >80 tests per hour in batch mode, 60 tests per hour with mixed load list configurations.
No takes yet. Share an insight, caveat, or question.
Wilson et al. (1998) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: