A comprehensive characterization of a Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometer utilizing a 1 tesla magnet with an electrospray ionization (ESI) source is presented. A resolution of up to 10,000,000 and a ppb-level mass accuracy has been achieved for atomic ions and light molecules produced with electron ionization. It has also been demonstrated that a resolution of up to 1,000,000 can be achieved for signals from biomolecule ions obtained with an ESI source, with the highest molecular weight for which the 13 C isotopic pattern was completely resolved being that of ubiquitin (8565 Da). Several features of FT-ICR mass spectrometry at low magnetic field are discussed and approaches to overcome the low-field constraints are presented.