Key result
Computer simulation of artificial heart implantation using a 3D thoracic model achieved contour extraction with a diameter error of approximately 2 mm (SD <0.5 mm) compared to a phantom.
Computer simulation of artificial heart implantation using 3D thoracic models from CT images is feasible with acceptable accuracy for assessing anatomic compatibility.
Supports simulation feasibility for artificial heart planning; leaves open clinical validation and human anatomic compatibility.
To assess the anatomic compatibility of an artificial heart (AH), we attempted to develop a computer environment that would facilitate a reliable simulation of an AH implanted in the human thorax. A three-dimensional thoracic computer model with a ventricle-resected heart was constructed, by using manually extracted contour points of the aorta, pulmonary artery, atria, atrioventricular valves, diaphragm, and thoracic wall from a set of consecutive CT images. Such a model enabled simulation of an AH implantation by orienting the AH model in it. Error evaluation on CT imaging and contour extraction with a Plexiglas cylindrical phantom showed that the diameter of the extracted phantom contour was approximately 2 mm smaller than its original with a standard deviation of <0.5 mm. Errors in contour and surface reconstruction could be reduced to far less than 1 mm under constrained conditions. A study on the influence of breathing revealed that variations in some thoracic dimensions between inspiration and expiration could reach 10 mm. In summary, computer simulation of AH implantation is a worthwhile approach with acceptable accuracy, although further considerations of extreme thoracic situations will be required.
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2000) studied Anatomic compatibility of an implantable artificial heart. Computer simulation of artificial heart implantation using a 3D thoracic model was evaluated on Error evaluation on CT imaging and contour extraction. Computer simulation of artificial heart implantation using a 3D thoracic model achieved contour extraction with a diameter error of approximately 2 mm (SD <0.5 mm) compared to a phantom.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: