Key result
In 9 test conditions at 4 L/min flow, membrane oxygenators slightly affected pulsatility, whereas the arterial pipe line caused large damping and head losses.
Arterial pipe lines in extracorporeal circuits are responsible for significant damping of pulsatile flow, suggesting that shorter lines are needed to optimize pulsatile cardiopulmonary bypass.
Answers are grounded in this paper and cite their sources.
May warrant shorter arterial tubing to minimize damping in pulsatile CPB; hypothesis-generating pending human validation.
An experimental analysis was carried out to evaluate the effects induced by two typical extracorporeal circuits on the pressure and flow generated by a roller pump with a pulsatile module. The hydraulic behaviour of the patient was simulated by means of a mechanical mock-up system consisting of a few lumped parameters reproducing the physiologic vascular impedance. Pressure and flow tracings were acquired at different locations along the circuit using an automatic data acquisition system. Nine lest conditions with different pulse frequency and systolic time values were examined using a mean volumetric flow rate of approximately 4 l/min. A complete analysis of the results obtained in terms of pressure drops and inflow-outflow differences across the components of the arterial line, as well as the calculation of the hydraulic pulsatile power along the circuit, allowed us to assess the influence of the various components upon the pulsatility. The results indicated that the membrane oxygenators tested slightly affect the pulsatility of the flow and the pressure; on the contrary the arterial pipe line is responsible for large damping and head losses. To optimize the use of pulsatile flow for cardiopulmonary bypass it is necessary to reduce the length of the arterial pipe lines thus integrating the circuit as much as possible.
No takes yet. Share an insight, caveat, or question.
Inzoli et al. (1997) studied Cardiopulmonary bypass (simulated). Membrane oxygenators and extracorporeal circuits was evaluated on Pressure drops and inflow-outflow differences across the components of the arterial line. In 9 test conditions at 4 L/min flow, membrane oxygenators slightly affected pulsatility, whereas the arterial pipe line caused large damping and head losses.
Synapse has enriched 10 related papers. Consider them for comparative context: