Key result
ECG-synchronized pulsatile ECLS maintained coronary and carotid artery diastolic velocities better than conventional nonpulsatile ECLS over 24 hours in an adult swine model.
Why the study?
Does ECG-synchronized pulsatile ECLS improve coronary and carotid blood flow velocities compared to nonpulsatile ECLS in an adult swine model?
RCT (n=9)
Randomly separated
Does ECG-synchronized pulsatile ECLS improve coronary and carotid blood flow velocities compared to nonpulsatile ECLS in an adult swine model?
Answers are grounded in this paper and cite their sources.
ECG-synchronized pulsatile ECLS appears to better maintain coronary and carotid artery diastolic velocities compared to conventional nonpulsatile ECLS in a swine model.
The objective of this study was to compare the effects of nonpulsatile and ECG-synchronized pulsatile extracorporeal life support on coronary and carotid blood flow velocities using transthoracic echocardiography and vascular ultrasound, respectively. Nine adult swine were randomly separated into nonpulsatile (NP, n = 5) and pulsatile (P, N = 4) groups and placed on ECLS for 24 h using an i-cor ECLS system. Noninvasive transthoracic images of the left and right coronary artery and the left carotid artery were acquired at the pre-ECLS (baseline), 30 min, 3, 6, 9, 12, and 24 h on-ECLS stages. The mean diastolic velocity of the left and right coronary arteries in the NP group significantly decreased after 24 h on ECLS compared to the baseline and 30 min ECLS stages (P < 0.05). There was no statistical difference in the mean diastolic velocity of the coronary arteries in the P group at 30 min, 3-, 6-, 9-, 12-, and 24-h ECLS compared to baseline. The P group showed a smaller decrease in the mean diastolic velocity of coronary arteries between the 30-min ECLS and 3-, 6-, 9-, 13-, 24-h ECLS stages compared to the NP group. The diastolic velocity of the left carotid artery in the NP group significantly decreased during 24-h ECLS compared to the P group (P < 0.05). An ECG-synchronized pulsatile ECLS system appeared to maintain coronary and carotid artery diastolic velocities better than conventional nonpulsatile ECLS. Further investigation of the perfusion modes during ECLS is warranted.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2018) conducted an RCT in Adult ECLS swine model (n=9). ECG-synchronized pulsatile extracorporeal life support (ECLS) vs. Conventional nonpulsatile ECLS was evaluated on Coronary and carotid blood flow velocities (p=<0.05). ECG-synchronized pulsatile ECLS maintained coronary and carotid artery diastolic velocities better than conventional nonpulsatile ECLS over 24 hours in an adult swine model.