Key result
The presence of a superficial vein across the clavicle predicted venous obstruction in patients with cardiovascular implantable electronic devices with a sensitivity of 96.3% and specificity of 77.5%.
Why the study?
Does the presence of a superficial vein across the clavicle predict venous obstruction in patients with cardiovascular implantable electronic devices?
Cross-Sectional (n=107)
No
Does the presence of a superficial vein across the clavicle predict venous obstruction in patients with cardiovascular implantable electronic devices?
Answers are grounded in this paper and cite their sources.
The presence of a superficial vein across the clavicle is a highly sensitive, non-invasive clinical sign for predicting venous obstruction in patients with cardiovascular implantable electronic devices.
BACKGROUND: Obstruction of the access vein is a well-known complication after cardiovascular implantable electronic device (CIED) implantation. In that case, well-developed collateral superficial veins are frequently observed on the skin surface around the CIED. The aim of this study was to clarify the relationship between venous obstruction and development of a superficial vein across the clavicle. METHODS AND RESULTS: A total of 107 patients scheduled for generator replacement, device upgrade, or lead extraction were enrolled. The skin surface around the device was photographed. A 20-ml bolus of contrast medium was injected into a peripheral arm vein on the side of CIED implantation, and contrast venography was performed. Venous obstruction was defined as luminal diameter narrowing >75%. Venography showed venous obstruction in 27 patients (25.2%). There were no statistically significant differences in patient characteristics between the venous obstruction and no venous obstruction group. Of 107 patients, 44 (41.1%) had a superficial vein across the clavicle on the side of CIED implantation. The sensitivity of the presence of a superficial vein across the clavicle in the diagnosis of venous obstruction was 96.3% and specificity was 77.5% (P<0.001). CONCLUSIONS: The presence of a superficial vein across the clavicle is useful for the prediction of venous obstruction in patients with CIED.
No takes yet. Share an insight, caveat, or question.
Hosoda et al. (2014) conducted a cross-sectional in Cardiovascular Implantable Electronic Device (CIED) implantation (n=107). Presence of a superficial vein across the clavicle vs. Absence of a superficial vein across the clavicle was evaluated on Venous obstruction (luminal diameter narrowing >75%) (p=<0.001). The presence of a superficial vein across the clavicle predicted venous obstruction in patients with cardiovascular implantable electronic devices with a sensitivity of 96.3% and specificity of 77.5%.
Synapse has enriched 10 related papers. Consider them for comparative context: