Key result
Device replacement via a subcostal approach without sternotomy successfully managed a Novacor LVAD malfunction, leading to uneventful recovery and successful transplantation one month later.
Case Report (n=1)
Novacor LVAD malfunction can be successfully managed with device replacement via a subcostal approach, avoiding the need for repeat sternotomy.
Answers are grounded in this paper and cite their sources.
May offer alternative to sternotomy in select LVAD cases; leaves open whether approach improves outcomes in larger series.
Implantable left-ventricular assist devices have been remarkably free of mechanical failures. We describe an uncommon Novacor N100 PCq LVAS malfunction caused by an internal short circuit of the device due to urine aspiration via the vent line. Device replacement was managed via a subcostal approach without sternotomy. Patient recovery was uneventful and successful transplantation was performed one month after the device exchange.
No takes yet. Share an insight, caveat, or question.
Hammel et al. (1998) conducted a case report in Novacor LVAD malfunction (n=1). Device replacement via a subcostal approach without sternotomy was evaluated on Patient recovery and successful transplantation. Device replacement via a subcostal approach without sternotomy successfully managed a Novacor LVAD malfunction, leading to uneventful recovery and successful transplantation one month later.