March 28, 2025EP EuropaceOpen Access

Ablation characteristics associated with steam pops in ablation index-guided radiofrequency ablation of atrial fibrillation

View Full Paper
Ask AI
Bookmark
Share

Key result

Impedance drop was significantly higher at steam pop sites compared to adjacent anatomical locations during radiofrequency ablation of atrial fibrillation (18.2 ± 9.5 Ω vs. 13.5 ± 4.8 Ω, P < 0.001).

Why the study?

Steam pops are a significant concern during radiofrequency ablation of AF, prompting investigation into their occurrence, incidence, and potential predictor ablation characteristics.

What are the incidence and predictors of steam pops during radiofrequency ablation in patients with atrial fibrillation?

Population

3263 patients with AF undergoing RF ablation

Comparison

Ablation sites with steam pops vs adjacent anatomical locations

Design

Observational study

Authors

XWXuesi WangCapital Medical UniversityZWZe‐Yang WuHebei North UniversityQWQi WuNingxia Medical University

Discussion

Loading...

Member takes

Implication

Supports impedance monitoring to anticipate steam pops in AF ablation; hypothesis-generating pending prospective validation.

Study Design

Type

Observational (n=3,263)

Structured PICO

What are the incidence and predictors of steam pops during radiofrequency ablation in patients with atrial fibrillation?

P
Population
3,263 patients with atrial fibrillation undergoing radiofrequency ablation.
E
Exposure
Ablation index-guided radiofrequency ablation (bilateral circumferential pulmonary vein ablation for paroxysmal AF, additional linear ablation for persistent AF)
C
Comparator
Ablation parameters at adjacent anatomical locations without steam pops
O
Outcome
Occurrence and potential predictors of steam popssafety

Main Result

Absolute Event Rate: 18.2% vs 13.5%

p-value: p=< 0.001

Steam pops during AF ablation are infrequent (2.5%), and a greater impedance drop is a significant predictor of their occurrence.

Cite This Study

Wang et al. (2025) conducted an observational in Atrial fibrillation (n=3,263). Steam pop sites vs. Adjacent anatomical locations was evaluated on Impedance drop (p=< 0.001). Impedance drop was significantly higher at steam pop sites compared to adjacent anatomical locations during radiofrequency ablation of atrial fibrillation (18.2 ± 9.5 Ω vs. 13.5 ± 4.8 Ω, P < 0.001).

synapsesocial.com/papers/6a9783e6275b4863ba2d1c8chttps://doi.org/10.1093/europace/euaf084

Topics

Atrial fibrillationPersistent AF managementAF ablation outcomes
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Advances in Irrigated Tip Catheter Technology for Treatment of Cardiac Arrhythmias2013 · 10 citations
  2. 2Radiofrequency ablation: technological trends, challenges, and opportunities2020 · 81 citations
  3. 3Prospective randomized comparison between a fixed ‘2C3L’ approach vs. stepwise approach for catheter ablation of persistent atrial fibrillation2015 · 139 citations
  4. 4Improving procedural and one-year outcome after contact force-guided pulmonary vein isolation: the role of interlesion distance, ablation index, and contact force variability in the ‘CLOSE’-protocol2018 · 366 citations
  5. 5Electro-characteristics of Myocardial Pouches and Reduction of the Frequency of Steam Pops During Radiofrequency Ablation2022 · 4 citations