Key result
Modified sub-selector sheath combined with a locking stylet successfully extracts fibrosed coronary sinus lead.
Case Report (n=1)
No
Fibrous adhesions can develop at the distal tip of coronary sinus leads as early as 4 months post-implantation, and modifying a standard sub-selector delivery sheath can enable safe transvenous extraction when manual traction and standard mechanical dilatation fail.
Answers are grounded in this paper and cite their sources.
May facilitate bailout extraction of early fibrosed CS leads; hypothesis-generating and leaves open need for validation before adoption.
An 84-year-old woman implanted with cardiac resynchronization therapy defibrillator underwent transvenous lead extraction 4 mo after the implant due to pocket infection. Atrial and right ventricular leads were easily extracted, while the attempt to remove the coronary sinus (CS) lead was unsuccessful. A few weeks later a new extraction procedure was performed in our center. A stepwise approach was used. Firstly, manual traction was unsuccessfully attempted, even with proper-sized locking stylet. Secondly, mechanical dilatation was used with a single inner sheath placed close to the CS ostium. Finally, a modified sub-selector sheath was successfully advanced over the electrode until it was free of the binding tissue. The post-extraction lead examination showed an unexpected fibrosis around the tip. No complications occurred during the postoperative course. Fibrous adhesions could be found in CS leads recently implanted requiring non-standard techniques for its transvenous extraction.
No takes yet. Share an insight, caveat, or question.
Bontempi et al. (2017) conducted a case report in Cardiac resynchronization therapy defibrillator pocket infection with Staphylococcus Pseudintermedius requiring coronary sinus lead extraction (n=1). Stepwise transvenous coronary sinus lead extraction using a locking stylet, mechanical dilatation, and a modified Attain Select sub-selector sheath cut 1.5 cm from the distal tip was evaluated on Successful transvenous extraction of fibrosed coronary sinus lead without complications. A modified sub-selector delivery sheath cut 1.5 cm from the distal part combined with a locking stylet successfully extracted a fibrosed coronary sinus lead 4 months post-implant without complications.
Synapse has enriched 10 related papers. Consider them for comparative context: