Why the study?
Radiation therapy can cause life-threatening malfunctions in cardiac implantable electronic devices due to electromagnetic interference or overcurrent. Previous reports had not described radiation therapy in patients implanted with an Aveir VR leadless pacemaker after removing a conventional transvenous pacemaker generator situated near pulmonary tumors.
Does switching to a leadless pacemaker and removing the transvenous pacemaker generator prevent radiation-induced device malfunction in lung cancer patients undergoing thoracic radiotherapy?
Case Report (n=2)
Does switching to a leadless pacemaker and removing the transvenous pacemaker generator prevent radiation-induced device malfunction in lung cancer patients undergoing thoracic radiotherapy?
Removing a conventional pacemaker generator and implanting an Aveir VR leadless pacemaker outside the radiation field is a feasible and safe strategy to prevent radiation-induced device malfunction in lung cancer patients requiring thoracic radiotherapy.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for leadless pacemaker strategy in thoracic radiotherapy; larger prospective studies needed before adoption.
Radiation therapy can cause a malfunction of cardiac implantable electronic devices (CIEDs) due to electromagnetic interference (EMI) or an overcurrent. Although rare, such malfunctions may lead to life-threatening outcomes. One reported case described inappropriate ventricular pacing caused by software errors in an implantable cardioverter defibrillator (ICD), which led to ventricular tachycardia1. Another case reported rapid atrial pacing due to pacemaker software errors, resulting in hypotension2.
No takes yet. Share an insight, caveat, or question.
Kumazawa et al. (2025) conducted a case report in Lung cancer and bradyarrhythmia (complete atrioventricular block or sick sinus syndrome) with previously implanted transvenous pacemaker (n=2). Implantation of a leadless pacemaker (Aveir VR) and removal of the existing transvenous pacemaker generator (with lead retention) was evaluated on Device malfunction or electrical parameter deterioration during and after radiation therapy (60 Gy/30 fractions).
Synapse has enriched 10 related papers. Consider them for comparative context: