Answers are grounded in this paper and cite their sources.
We suggest a new explanation for the lack of auditory response to electric stimulation in children with cochlear implants: the very narrow internal auditory canal, 1 to 2 mm in diameter, and the probable absence of the cochlear nerve. This defect can be seen on high-resolution computed tomographic x-ray studies and may represent aplasia of the auditory-vestibular nerve. We report on eight children with this anomaly, three of whom have received implants and failed to respond with a sensation of sound. Identification of this problem on screening x-ray films is a contraindication to cochlear implantation for auditory stimulation.
No takes yet. Share an insight, caveat, or question.
Shelton et al. (1989) studied this question.