Key result
The advanced hybrid closed-loop system resulted in similar early postprandial glycemic control compared to the 670G system after adjusting for premeal glucose levels (adjusted mean difference 1.8 mg/dL, p=0.10).
Why the study?
It was unknown whether the reduced hyperglycemia observed with an advanced hybrid closed-loop system compared to a first-generation system was specifically related to better mealtime glycemic control.
Does the Medtronic advanced hybrid closed-loop system improve postprandial glycemic control compared to the 670G system in young patients with type 1 diabetes?
RCT (n=112)
Open-label
Randomized crossover
Yes
Does the Medtronic advanced hybrid closed-loop system improve postprandial glycemic control compared to the 670G system in young patients with type 1 diabetes?
Mean Difference: 1.8 (95% CI -0.4–4)
Absolute Event Rate: 177% vs 180%
p-value: p=0.10
The reduced overall hyperglycemia seen with the advanced hybrid closed-loop system compared to the 670G system is likely driven by later postprandial or overnight improvements rather than early postprandial control.
Early postprandial similarity implies overall benefits arise later or overnight; extends timing-specific understanding of closed-loop systems.
Background: We recently reported that use of an “advanced” hybrid closed-loop system reduced hyperglycemia without increasing hypoglycemia compared to a first-generation system. The aim of this analysis was to evaluate whether this improved performance was specifically related to better mealtime glycemic control. Methods: We conducted a secondary analysis of postprandial glycemic control in an open-label, multinational, randomized crossover trial of 112 participants with type 1 diabetes, aged 14–29, of the Medtronic MiniMed™ 670G hybrid closed-loop system (670G) versus the Medtronic advanced hybrid closed-loop (AHCL) system, for 12 weeks each. We compared glycemic and insulin delivery metrics over a 3 h horizon across all meals to assess system performance and outcomes. Results: Overall meal size and premeal insulin on board were similar during run-in and between 670G and AHCL arms. Compared with 670G arm, premeal, peak, and mean glucose levels were numerically lower in the AHCL arm (167 ± 23, 231 ± 23, and 177 ± 20 mg/dL vs. 175 ± 23, 235 ± 23, and 180 ± 19 mg/dL, respectively), with a trend to lower hyperglycemia level 2 in AHCL arm. Adjusting for premeal glucose level, all postmeal outcomes between 670G and AHCL were statistically similar. Prandial insulin delivery also was similar in both treatment arms (21 ± 9 vs. 23 ± 10 U), with a shift in basal/bolus ratio from 28%/71% in 670G arm to 20%/80% in AHCL arm. Conclusions: Reduced hyperglycemia with AHCL compared to 670G was not related to early postprandial glycemic excursions after adjusting for premeal glucose level (<3 h after meal), but likely to later (>3 h) postprandial or overnight improvements. Further refinements to mealtime bolus algorithms and strategies may more optimally control prandial glycemic excursions.
No takes yet. Share an insight, caveat, or question.
Weinzimer et al. (2022) conducted an RCT in Type 1 diabetes (n=112). Medtronic advanced hybrid closed-loop (AHCL) system vs. Medtronic MiniMed 670G hybrid closed-loop system was evaluated on Mean postprandial glucose level (mg/dL) (MD 1.8, 95% CI -0.4 to 4.0, p=0.10). The advanced hybrid closed-loop system resulted in similar early postprandial glycemic control compared to the 670G system after adjusting for premeal glucose levels (adjusted mean difference 1.8 mg/dL, p=0.10).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: