Answers are grounded in this paper and cite their sources.
We report a novel lateral flow chromatography strip (LFCS) that integrates a target-responsive triplex DNA molecular switch with a carbon nanotube (CNT) label for the point-of-care (POC), ultrasensitive detection of thrombin. The biosensor employs a triplex DNA structure, formed between a thrombin-specific aptamer strand (P0) and an auxiliary strand (P1) covalently anchored to carboxylated CNTs. In the absence of thrombin, the intact triplex sterically hinders P1, preventing signal generation. Upon target binding, the triplex disassembles, activating P1 to hybridize with a capture probe on the test line and leading to the accumulation of CNTs as a visually quantifiable black band. This design leverages the high surface area of CNTs and preserves the aptamer's native affinity. The assay achieves a detection limit of 0.15 nM in buffer, representing a 16.7-fold improvement in sensitivity over a conventional aptamer-gold nanoparticle LFCS. It demonstrates high specificity against non-target proteins and reliable performance in a 10% human serum matrix. By combining a specific conformational switch with a robust nanomaterial label, this work provides a versatile and powerful platform for POC protein detection with significant potential in clinical diagnostics.
No takes yet. Share an insight, caveat, or question.
Cheng et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: