Key result
Non-contact subpixel image registration accurately quantified jugular venous displacement waveforms from skin displacements, achieving an average correlation score of 0.93 ± 0.05 compared to a laser displacement sensor.
Why the study?
Existing image-based methods for measuring jugular venous waveforms cannot directly measure skin deformations, rely on photoplethysmography, and have limited accuracy and robustness for clinical use.
Does a non-contact camera-based subpixel image registration method accurately measure jugular venous pulse waveforms compared to a laser displacement sensor in healthy subjects?
Observational (n=19)
No
Does a non-contact camera-based subpixel image registration method accurately measure jugular venous pulse waveforms compared to a laser displacement sensor in healthy subjects?
Effect estimate: r = 0.93 ± 0.05
A novel non-contact camera-based method using subpixel image registration can accurately measure jugular venous pulse waveforms, showing high correlation with laser displacement sensors.
May facilitate non-contact JVP monitoring; hypothesis-generating before clinical validation in cardiovascular patients.
The jugular venous (JV) pressure waveform is a non-invasive, proven indicator of cardiovascular disease. Conventional clinical methods for assessing these waveforms are often overlooked because they require specialised expertise, and are invasive and expensive to implement. Recently, image-based methods have been used to quantify JV pulsation waveforms on the skin as an indirect way of estimating the pressure waveforms. However, these existing image-based methods cannot explicitly measure skin deformations and rely on the use of photoplethysmography (PPG) devices for identification of the pulsatile waveforms. As a result, they often have limited accuracy and robustness and are unsuitable in the clinical environment. Here, we propose a technique to directly measure skin deformations caused by the JV pulse using a very accurate subpixel registration algorithm. The method simply requires images obtained from the subject's neck using a commodity camera. The results show that our measured waveforms contained all of the essential features of diagnostic JV waveforms in all of 19 healthy subjects tested in this study, indicating a significantly important capability for a potential future diagnostic device. The shape of our measured JV displacement waveforms was validated using waveforms measured with a laser displacement sensor, where the average correlation score between the two waveforms was 0.93 ± 0.05. In addition, synchronously recorded ECG signals were used to verify the timings of diagnostic features of the measured waveforms. To our knowledge, this is the first use of image registration for direct measurement of JV displacement waveforms. Significant advantages of our novel method include the high precision of our measurements, and the ability to use ordinary cameras, such as those in modern mobile phones. These advantages will enable the development of affordable and accessible devices to measure JV waveforms for cardiac diagnostics in the clinical environment. Future devices based on this technology may provide viable options for telemedicine applications, point of care diagnostics, and mobile-based cardiac health monitoring systems.
No takes yet. Share an insight, caveat, or question.
Tang et al. (2018) conducted an observational in Healthy (n=19). Non-contact subpixel image registration (P-SG-GC algorithm) vs. Laser displacement sensor was evaluated on Correlation score between camera-measured and laser-measured JV displacement waveforms (r = 0.93 ± 0.05). Non-contact subpixel image registration accurately quantified jugular venous displacement waveforms from skin displacements, achieving an average correlation score of 0.93 ± 0.05 compared to a laser displacement sensor.