Key result
A novel modified mouthguard delivering oxygen and sampling carbon dioxide was successfully used in over 250 upper endoscopy cases with minimal complications, avoiding the risks of oesophageal occlusion.
Answers are grounded in this paper and cite their sources.
A novel modified mouthguard may provide a safer alternative to gastro-laryngeal tubes for maintaining airway patency and reliable capnography during deep sedation for ERCP.
We read with interest the article by Dr Gaitini and colleagues describing the use of gastro-laryngeal tubes during anaesthesia for endoscopic retrograde cholangiopancreatography (ERCP) [1]. Optimal airway management during ERCP procedures is open to debate, and we acknowledge the authors’ efforts to develop a supra-glottic technique that maintains airway patency, protects against gastric regurgitation and enables assisted ventilation during apnoeic episodes. However, experience with laryngeal tubes (and their variants) in the UK is limited and there remains considerable uncertainty about the benefits of oesophageal occlusion vs the risks of pharyngeal and oesophageal trauma. There has been at least one case report of oesophageal rupture with the use of a Combitube [2]. Propofol sedation has been used extensively for endoscopic procedures. Coté et al. [3] prospectively studied 799 patients undergoing ERCP, endoscopic ultrasound and small bowel enteroscopy procedures with anaesthetic nurse-administered propofol. No patients required tracheal intubation during the procedure, suggesting no episodes of significant aspiration. Rex et al. [4] conducted a worldwide safety review and collected information retrospectively on 646 080 upper and lower endoscopic procedures with non-anaesthetists administering propofol sedation. They reported eleven rescue tracheal intubations overall. Our experience across two hospitals during a 10-year period (approximately 3000 upper endoscopy procedures) is also that the risk of regurgitation and aspiration is very low. Our patients are rigorously fasted and if they have risk factors for regurgitation (particularly gastric outlet obstruction) we perform elective tracheal intubation. On insertion of the endoscope, we find that the stomach is nearly always empty. If there are residual gastric contents they can usually be suctioned by the operator, and if not, the endoscope is removed and we perform tracheal intubation following rapid sequence induction of anaesthesia. Finally, upper endoscopic procedures are performed in the semi-prone or prone position, providing a degree of protection in the event of regurgitation. Given the low aspiration risk, the most common worldwide approach to these patients is to maintain an open airway, with a mouthguard in-situ that provides access and protection for the endoscope. We favour this strategy, although we have encountered some difficulties ensuring adequate supplemental oxygen delivery and reliable capnography monitoring. In particular, capnography is widely accepted to improve safety during deep sedation [5]. Absence of capnography may have contributed to the relatively high incidence of hypoxaemia presented by Gaitini et al. We have recently developed a novel mouthguard that delivers oxygen and samples carbon dioxide simultaneously from the nose and mouth, using two nasal prongs and two oral channels (Fig. 1). It is fitted with the patient fully conscious in the semi-prone position. Deep sedation or light general anaesthesia is induced and maintained with propofol boluses in accordance with a local protocol. With the mouthguard in place, capnography monitoring is highly reliable. This facilitates the optimal titration of propofol to achieve patient comfort whereas it avoids apnoeic episodes. We have successfully used this new device in over 250 cases with minimal complications. We are now carrying out formal evaluation studies. Pennine Mouthguard (Pennine Healthcare Ltd, Derby, UK). Port A; oxygen attachment, Port B; capnography attachment, Face L; position of lips, Face T; position of teeth. We believe there is no evidence to support the use of airway devices that occlude the oesophagus during advanced upper endoscopy. In fact, the use of gastro-laryngeal tubes may be dangerous, and much larger studies would be necessary to demonstrate safety. We feel that appropriate patient selection, good teamwork with the gastroenterologists and the use of a modified mouthguard are the keys to safety and success. The modified mouthguard was designed by Dr Keeling and manufactured by Pennine Healthcare. No external funding has been received and none of the authors stand to gain financially from commercial sales. Previously posted at the Anaesthesia Correspondence website: http://www.anaesthesiacorrespondence.com.
No takes yet. Share an insight, caveat, or question.
Toner et al. (2011) conducted a letter in Endoscopic retrograde cholangiopancreatography (ERCP) and advanced upper endoscopy (n=250). Modified mouthguard (Pennine Mouthguard) vs. Gastro-laryngeal tubes (historical/discussed) was evaluated on Complications. A novel modified mouthguard delivering oxygen and sampling carbon dioxide was successfully used in over 250 upper endoscopy cases with minimal complications, avoiding the risks of oesophageal occlusion.
Synapse has enriched 10 related papers. Consider them for comparative context: