Answers are grounded in this paper and cite their sources.
Various sugars, peptides, and lipids were analyzed on a Fourier transform mass spectrometer using laser desorption and ionization with and without the assistance of matrixes. A compact Nd:YAG laser with an output at 1.06 microns corresponding to fundamental frequency was employed. Gram-negative and Gram-positive bacteria were also subjected to laser desorption mass spectrometry. Characteristics ions of conjugated lipid, formed by attachment of alkali metal cations, endogenous to the cells, were observed. Particle/liquid matrixes (e.g., cobalt in glycerol) proved to be useful with the 1.06-micron laser. The particles absorb efficiently laser radiation in a broad wavelength range. The liquid provides the same advantages as in fast atom bombardment: increased signal-to-noise ratios and enhanced sample lifetimes. The effect of laser power on total ion current was shown to differ for samples with and without the particle/liquid matrix. The Fourier transform analyzer provides MS/MS capability for both positive and negative ions from complex mixtures. Ions desorbed externally are introduced into the cell via a quadrupole ion guide with a lower mass cutoff. Such a setup allows matrix ions to be excluded and thus provides excellent signal-to-noise ratios for lower mass range fragment ions formed inside the cell.
No takes yet. Share an insight, caveat, or question.
Ho et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: