The vibrational and electromechanical characteristics of a wide range of 1-3 composite structures, comprising ceramic pillars aligned within a polymer phase, are considered using finite-element analysis. The influence of pillar geometry, ceramic volume fraction, and pillar orientation is described in terms of overall transduction efficiency. It is shown that the finite-element method provides a versatile means of analysis and the results obtained permit a set of useful design guidelines to be developed. In general, a small pillar aspect ratio and a relatively high volume fraction provides the most satisfactory performance, in terms of electromechanical efficiency and uniformity of thickness dilation.
No takes yet. Share an insight, caveat, or question.
Hossack et al. (1991) studied this question.