Answers are grounded in this paper and cite their sources.
We proposed a novel noninvasive power-supply system for medical implants at a deep position in the body. The system consists of the magnetic coupling and the small generator. The former system could keep the power of 1 W supplying until 2.5 cm from the skin and the maximum output power was 3.1 W at 1.5 cm. We have manufactured more powerful generators as the next step. The improvement points are to decrease or eliminate the interference between the magnets at the implanted rotor. As a result, the output power increases 6.6 W and the 1 W supplying distance is extended to 3.0 cm.
No takes yet. Share an insight, caveat, or question.
Suzuki et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: