Key result
Intracardiac electrocardiography-guided catheter tip positioning for TIVAD placement achieved an ideal position rate of 96.8%, comparable to 97.1% with X-ray guidance.
Why the study?
The study was conducted to evaluate the accuracy and safety of intracardiac electrocardiography compared with X-ray guidance for catheter tip positioning during totally implantable venous access device placement.
Does intracardiac electrocardiography (IC-ECG) guidance improve the ideal position rate and reduce complications compared to X-ray guidance in adult oncology patients undergoing TIVAD placement?
Observational (n=465)
No
Does intracardiac electrocardiography (IC-ECG) guidance improve the ideal position rate and reduce complications compared to X-ray guidance in adult oncology patients undergoing TIVAD placement?
IC-ECG is a safe and accurate alternative to X-ray guidance for TIVAD catheter tip positioning, offering comparable success rates with the potential to reduce radiation exposure.
Answers are grounded in this paper and cite their sources.
IC-ECG yields comparable tip accuracy and safety to X-ray in TIVAD placement; leaves open whether it can routinely replace radiography in practice.
Purpose: To assess the accuracy and safety of intracardiac electrocardiography (IC-ECG) in positioning catheter tips for Totally Implantable Venous Access Device (TIVAD) placement. Methods: This study conducts a retrospective analysis of patient data collected from The Affiliated Suzhou Hospital of Nanjing Medical University. Patients were categorized into two groups based on the method used for catheter tip positioning: the IC-ECG group and the X-ray group. Propensity Score Matching (PSM) is employed to balance the baseline characteristics of the two groups in a 1:2 ratio. The primary outcomes of interest are the ideal position rate of the TIVAD tip and the incidence of complications. Additionally, multivariable logistic regression analysis will be utilized to identify risk factors associated with complications. The area under the ROC curve (AUC) will be calculated to evaluate the diagnostic performance of the detection methods. Results: During the period from January 2023 to June 2024, a total of 493 adult oncology patients received TIVAD. After applying PSM, 465 patients were included in the retrospective analysis, comprising 221 males (47.5%) and 244 females (52.5%), with a mean age of 62.6 years. The ideal position rates for the catheter tips were 150 cases (96.8%) in IC-ECG group and 301 cases (97.1%) in X-ray group ( p = 0.834). The distance from the catheter tip to the carina was measured at 1.60 ± 0.356 and 1.41 ± 0.34 vertebral body units, respectively ( p < 0.01). The overall complication rates were comparable between the two groups. Multivariate logistic regression analysis indicated that BMI is an independent risk factor for TIVAD-related complications (OR = 0.437, 95% CI: 0.319–0.563, p < 0.001). The ROC curve analysis revealed that the area under the curve (AUC) for BMI was 0.926 (95% CI: 0.886–0.926, p < 0.001). Conclusion: Intracardiac electrocardiography (IC-ECG)-guided catheter tip positioning for TIVAD placement has demonstrated accuracy, feasibility, and safety. Given its potential to reduce radiation exposure, improve safety, and offer cost-effectiveness, IC-ECG presents a promising method for catheter tip positioning in TIVAD procedures. Based on these advantages, IC-ECG could be considered a preferable method in clinical practice.
No takes yet. Share an insight, caveat, or question.
Zhao et al. (2025) conducted an observational in Oncology patients requiring TIVAD (n=465). Intracardiac electrocardiography (IC-ECG) guidance vs. X-ray guidance was evaluated on Ideal position rate of the TIVAD tip (p=0.834). Intracardiac electrocardiography-guided catheter tip positioning for TIVAD placement achieved an ideal position rate of 96.8%, comparable to 97.1% with X-ray guidance.