Key result
The intracavitary ECG method for tunneled dialysis catheter insertion without fluoroscopy achieved adequate flow in 139 of 142 cases and a 98.5% correct matching rate with chest PA view.
Why the study?
Does intracavitary ECG monitoring with ultrasound guidance allow for safe and effective insertion of tunneled dialysis catheters without fluoroscopy in adult hemodialysis patients?
Observational (n=142)
No
Does intracavitary ECG monitoring with ultrasound guidance allow for safe and effective insertion of tunneled dialysis catheters without fluoroscopy in adult hemodialysis patients?
Intracavitary ECG combined with ultrasound guidance is a safe and feasible method for bedside insertion of tunneled dialysis catheters without fluoroscopy in patients with an evident P wave.
Answers are grounded in this paper and cite their sources.
May support fluoroscopy-free tunneled catheter insertion in select patients with evident P waves; hypothesis-generating and requires randomized validation before practice change.
PURPOSE: Recently, there have been many reports that exact central vein catheter tip positioning was possible using intracardiac electrocardiographic (ECG) monitoring. Ultrasonic guidance in combination with intracardiac ECG monitoring may allow for a tunneled dialysis catheter to be inserted at the bedside without using fluoroscopy. Therefore we report on the intracavitary ECG method for insertion of a tunneled dialysis catheter with ultrasound guidance and the feasibility, safety, effectiveness, complications and limitations of this method. METHODS: From April 2012 to June 2014, we evaluated 142 hemodialysis (HD) patients who were dialyzed by a tunneled dialysis catheter that was inserted using intracardiac ECG monitoring without fluoroscopic usage. We checked the intracardiac P wave and the point at which it gradually rose to the highest P wave morphology, we stopped inserting the catheter. RESULTS: Catheter flow during dialysis was adequate in 139 cases. There were three cases where it malfunctioned. Catheter malposition occurred in 6 out of 142 cases. The correct matching rate between the intracardiac ECG and chest posteroanterior (PA) view was 98.5%. No significant complications developed. CONCLUSIONS: To conclude, in this single-center study, the intracavitary ECG method for assessing the position of the tip of tunneled dialysis catheter was proven to be safe and feasible in virtually all adult patients who had an evident P wave at the basal ECG tracking.
No takes yet. Share an insight, caveat, or question.
Cho et al. (2015) conducted an observational in Hemodialysis (n=142). Intracavitary ECG method for insertion of a tunneled dialysis catheter was evaluated on Correct matching rate between the intracardiac ECG and chest posteroanterior (PA) view. The intracavitary ECG method for tunneled dialysis catheter insertion without fluoroscopy achieved adequate flow in 139 of 142 cases and a 98.5% correct matching rate with chest PA view.
Synapse has enriched 10 related papers. Consider them for comparative context: