Key result
Six patients with severe access problems underwent implantation of a leadless pacemaker system (Micra) in Greece.
Case Report (n=6)
No
Answers are grounded in this paper and cite their sources.
This initial case series describes the early experience and procedural technique of leadless pacemaker implantation in Greece for patients with severe venous access issues.
cardiac pacing; pacemaker; leadless pacemaker; bradycardia Leadless pacing systems have recently emerged as a reliable therapeutic alternative to conventional pacemakers in providing therapy for patients with bradyarrhythmias.1,2 Initial studies demonstrate favorable efficacy and safety results compared to transvenous pacemakers.3e8 We report the first six cases of a leadless pacemaker system (Micraä Transcatheter Pacing System, Medtronic, Minneapolis, MN, USA) implantation in Greece that took place in the electrophysiology laboratory of Hippokrateion General Hospital of Athens between April and November 2016.Because of the relatively high cost of leadless pacemakers compared to the conventional ones, the use of this technology is still limited in Greece, and currently such implantations are restricted mostly to patients who present with severe access problems, unlikely to be overcome by conventional transvenous implantation techniques.Data of our patients are presented in Table 1.Procedure: Implantation was performed under fluoroscopy, after obtaining informed consent from the patients, with local anesthesia.After introducing a 24-French sheath into the right femoral vein, a deflectable delivery catheter with the pacemaker adjusted on its distal part was advanced through the inferior vena cava and the right atrium to the right ventricle (RV).Subsequently, the outer sheath was retracted, allowing the device tines to be deployed, fixing the pacemaker in the right ventricular trabeculae.Adequate fixing of the system was confirmed mechanically by the "tug-test" while the pacemaker still Peer review under responsibility of Hellenic Society of Cardiology.
No takes yet. Share an insight, caveat, or question.
Sideris et al. (2017) conducted a case report in bradyarrhythmias (n=6). Leadless pacemaker system (Micra Transcatheter Pacing System) was evaluated. Six patients with severe access problems underwent implantation of a leadless pacemaker system (Micra) in Greece.
Synapse has enriched 10 related papers. Consider them for comparative context: