Key result
An epicardial left ventricular lead implanted during open-heart surgery was found fractured after two years, highlighting the necessity of rigorous lead evaluation at generator implant.
This case highlights the importance of good surgical implant technique and rigorous lead evaluation for signs of impending failure at generator implant.
Epicardial left ventricular leads can be implanted at open-heart surgery for cardiac resynchronization therapy. We report a 2-year-old fractured epicardial left ventricular lead detected at generator implant. It highlights the importance of good surgical implant technique and of rigorous lead evaluation for signs of impending failure at generator implant.
No takes yet. Share an insight, caveat, or question.
Larsen et al. (2020) studied this question. An epicardial left ventricular lead implanted during open-heart surgery was found fractured after two years, highlighting the necessity of rigorous lead evaluation at generator implant.