This is the authors' abstract. We don't add key points for this paper.
We realize a single particle microscope by using deterministically extracted laser-cooled ⁴⁰Ca⁺ ions from a Paul trap as probe particles for transmission imaging. We demonstrate focusing of the ions to a spot size of 5.8±1.0 nm and a minimum two-sample deviation of the beam position of 1.5 nm in the focal plane. The deterministic source, even when used in combination with an imperfect detector, gives rise to a fivefold increase in the signal-to-noise ratio as compared with conventional Poissonian sources. Gating of the detector signal by the extraction event suppresses dark counts by 6 orders of magnitude. We implement a Bayes experimental design approach to microscopy in order to maximize the gain in spatial information. We demonstrate this method by determining the position of a 1 μm circular hole structure to a precision of 2.7 nm using only 579 probe particles.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Jacob et al. (2016) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: