Key result
LVADs produce electromagnetic interference disrupting ICD communication, with Heartware only interacting with Biotronik ICDs at 0-6 cm, while HeartMate 3 interfered with all but Medtronic ICDs.
Why the study?
Electromagnetic interference produced by an LVAD can preclude ICD interrogation with external programmers, but the extent of this interaction has not been clarified.
Does electromagnetic interference from running LVADs affect the ability to interrogate and reprogram ICDs?
Does electromagnetic interference from running LVADs affect the ability to interrogate and reprogram ICDs?
HeartMate 3 LVADs produce significant electromagnetic interference that disrupts communication with most ICDs, a safety issue that can be mitigated by increasing the distance between the devices.
BACKGROUND: Implanted cardioverter defibrillators (ICDs) and left ventricular assist devices (LVADs) are established interventions that prolong life in advanced heart failure, but their combination has not been demonstrated as beneficial. Electromagnetic interference (EMI) produced by a LVAD can preclude ICD interrogation with external programmers. We undertook a systematic evaluation of the LVAD-ICD interaction "in-vitro" to clarify the extent of this interaction. METHODS: Using explanted ICDs and VADs in a mock physiological rig, we assessed interrogation and reprogramming of ICD devices in the presence of a running LVAD. When connectivity between the ICD programmer and the ICD failed, we attempted three different techniques to re-establish connectivity: (1) Electromagnetic shielding of the ICD with a pseudo-faraday cage; (2) altering the LVAD speed; and (3) increasing the distance between the VAD and the ICD. RESULTS: We tested a total of 24 ICDs from different manufacturers in the presence of the Heartware (HW) and HeartMate 3 (HM3) LVADs. With HW, we only observed interaction with Biotronik ICD devices at very close range (0-6 cm). With HM3, only Medtronic ICD devices showed no interaction. Interactions could be mitigated by increasing the VAD-ICD distance. CONCLUSIONS: LVADs, notably the HM3, produce EMI that interferes with the communication between an ICD and its respective programmer. This may need to be considered when choosing the type of VAD to implant in patients with a previously implanted left-sided ICD. The only safe way to regain connectivity is to increase the distance between the VAD and the ICD, with patients raising their arm above their head.
No takes yet. Share an insight, caveat, or question.
Schnegg et al. (2022) studied Advanced heart failure (n=24). Left ventricular assist device (LVAD) operation was evaluated on Interrogation and reprogramming of ICD devices in the presence of a running LVAD. LVADs produce electromagnetic interference disrupting ICD communication, with Heartware only interacting with Biotronik ICDs at 0-6 cm, while HeartMate 3 interfered with all but Medtronic ICDs.