Answers are grounded in this paper and cite their sources.
A penetrameter, which is a device for checking the kVp calibration of an X-ray generator, has been designed and constructed in the University of Ilorin using locally available materials. The device consists of a 2 mm thick sheet of aluminium used as a beam hardener, stacks of various numbers of aluminium discs for beam attenuation, Perspex for support and an X-ray film cassette for detection. The identification of dots of equal optical density is made using a densitometer also made by one of the authors. The device has been tested in the Radiology department of the University of Ilorin Teaching Hospital on a diagnostic range X-ray machine and found to be satisfactory. It is simpler and less expensive than imported penetrameter cassettes.
No takes yet. Share an insight, caveat, or question.
Bassey et al. (1989) studied this question.