In this paper, a general method to calibrate the absorption coefficient of an absorber and the reduced scattering coefficient of a liquid diffusive medium, based on time-resolved measurements, is reported. An exhaustive analysis of the error sources affecting the estimation is also performed. The method has been applied with a state-of-the-art time-resolved instrumentation to determine the intrinsic absorption coefficient of Indian ink and the reduced scattering coefficient of Intralipid-20%, with a standard error smaller than 1% and 2%, respectively. Finally, the results have been compared to those retrieved for the same compounds by applying a continuous wave method recently published, obtaining an agreement within the error bars. This fact represents a cross validation of the two independent calibration methods.
No takes yet. Share an insight, caveat, or question.
Spinelli et al. (2007) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: