Key result
Trillium surface coating of oxygenators did not significantly affect the adhesion of activated platelets, white blood cells, and fibrin compared to non-coated oxygenators during CABG.
Why the study?
Does a Trillium-coated oxygenator reduce the adhesion of thrombotic components compared to a non-coated oxygenator in patients undergoing elective CABG?
RCT (n=20)
randomized
Does a Trillium-coated oxygenator reduce the adhesion of thrombotic components compared to a non-coated oxygenator in patients undergoing elective CABG?
Trillium coating on oxygenators does not significantly affect the primary adhesion of thrombotic components during CABG under full systemic heparinization.
Trillium coating provides no reduction in thrombotic adhesion during CABG; challenges presumed benefits of surface modification in heparinized circuits.
The Trillium coating is designed to minimize adsorption of protein and the attachment of cells and other particles. The present study was undertaken to investigate the effect of surface coating on the adhesion of thrombotic components (activated platelets, white blood cells and fibrin) to the surface of a clinically used oxygenator. Twenty patients undergoing elective coronary artery bypass grafting (CABG) were randomized to one of the two oxygenator groups: non-coated (NC, n = 10) or Trillium-coated (TC, n = 10). Platelet and white blood cell counts and factor XIIa concentrations were determined prior to the induction of anesthesia and at the end of cardiopulmonary bypass (CPB). Binding of activated platelets, white blood cells and fibrin to the artificial surfaces was quantified by means of antibody binding and histological validation was achieved by scanning electron microscopy. Patient demographic and CPB data were similar for the two groups. No significant differences between the groups were found for any of the tested thrombotic components. However, observations from our scanning electron microscopy suggested a release of formed particles from the Trillium-coated surface. Primary adhesion of activated platelets, white blood cells and fibrin to the artificial surface of the venous blood inlet from an oxygenator is not affected by the Trillium surface coating under conditions of full systemic heparinization.
No takes yet. Share an insight, caveat, or question.
Goor et al. (2006) conducted an RCT in elective coronary artery bypass grafting (CABG) (n=20). Trillium-coated oxygenator vs. non-coated oxygenator was evaluated on Adhesion of thrombotic components (activated platelets, white blood cells and fibrin) to the surface of the oxygenator. Trillium surface coating of oxygenators did not significantly affect the adhesion of activated platelets, white blood cells, and fibrin compared to non-coated oxygenators during CABG.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: