A method of measuring the proportion of primary radiation which traverses the patient's body is described. From such measurements the effective shape of the patient is found. The effective shape is the shape of the water equivalent body which would transmit the same proportion of the primary beam to the level of the point under consideration as the actual body dose. The calculation of dose is carried out on the assumption that the patient's body is a water-equivalent body whose shape is this effective shape. The accuracy of the calculation depends on the accuracy with which the effective shape can be drawn and, in cases in which an appreciable proportion of the total dose is delivered by scattered radiation, on the accuracy with which the scatter contribution in the patient's body can be determined. Examples of the application of the method to the estimation of doses in the thorax, skull and pelvis, are described.