A highly sensitive HPLC method was developed for the determination of xenoestrogenic compound, bisphenol A (BPA) in human breast milk samples. After a two-step liquid-liquid extraction, BPA was derivatized with fluorescent labeling reagent, 4-(4,5-diphenyl-1H-imidazol-2-yl)benzoyl chloride (DIB-Cl). The excess fluorescent reagent could be removed effectively using a column-switching system. The separation of DIB-BPA from endogenous materials in milk was carried out on two C(18) columns and fluorescence intensity was monitored at 475 nm with the excitation of 350 nm. A good linearity (r = 0.994) was observed of BPA in the concentration range of 0.2-5.0 ng mL(-1) in breast milk, and the detection limit was 0.11 ng mL(-1) at a signal-to-noise ratio of 3. Intra- and inter-day precision (RSD, %) were less than 8.7 and 10.4, respectively. Twenty-three breast milk samples of healthy lactating women were analyzed for the BPA concentration; the mean value was 0.61 +/- 0.20 ng mL(-1), with no correlation to the lipid content of milk samples.