Spectroscopic study demonstrates microsecond dual-comb spectroscopy in the 10–12.5 μm fingerprint region for chlorine monoxide, highlighting a tool for tracking transient radical kinetics.
Dual-comb spectroscopy enables broadband, high-resolution measurements with microsecond temporal resolution, but extending this capability to the 10–12.5 μm molecular fingerprint region remains technically challenging, particularly for transient radical kinetics. Here, we demonstrate microsecond-resolved dual-comb spectroscopy in this spectral range using electro-optic combs and difference-frequency generation in an orientation-patterned gallium phosphide (OP-GaP) crystal. Operation near a turning-point quasi-phase-matching condition at ∼140 °C reduces the wavelength sensitivity of the nonlinear conversion, enabling robust tuning of the idler comb over 83 cm−1, corresponding to ∼1.2 μm near 12 μm, by adjusting only the signal-comb center wavelength while keeping the pump wavelength and crystal temperature fixed. We further apply this platform to high-resolution, microsecond-resolved spectroscopy of transient chlorine monoxide (ClO) near 12 μm. Time-resolved dual-comb spectra capture the temporal evolution of ClO produced by the Cl + O3 reaction with a temporal resolution of 1.5 μs, enabling quantitative determination of the ClO formation rate coefficient. These results establish turning-point OP-GaP quasi-phase matching as an effective strategy for extending electro-optic dual-comb spectroscopy into the long-wavelength fingerprint region and validate this platform as a promising tool for quantitative, microsecond-resolved studies of short-lived radicals, particularly atmospherically relevant halogen oxides.
No takes yet. Share an insight, caveat, or question.
Luo et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: