Retrospective case series reveals rapid proficiency and high implant accuracy in patients undergoing robotic partial knee arthroplasty, indicating minimal learning curve barriers.
Purpose To evaluate the platform-specific learning curve and early tibial radiographic agreement of imageless ROSA ® -assisted medial unicompartmental knee arthroplasty (UKA), with short-term clinical findings assessed exploratorily. Methods This retrospective consecutive case series included 30 patients who underwent ROSA ® -assisted medial UKA with the Persona ® Partial Knee implant between January and April 2026 at a high-volume arthroplasty referral centre. A continuous learning-curve cumulative summation (LC-CUSUM) analysis of total surgical duration was performed against an independently defined surgeon-specific historical benchmark derived from institutional operative-register skin-to-skin times for manual UKA performed by the same surgeon (20 ± 4 min). Adequate performance was defined as a mean duration of 25 min, inadequate performance as 30 min, and the reference standard deviation as 4 min. Type I and type II error targets were 0.05 and 0.10, respectively; the decision boundary was calibrated by Monte Carlo simulation over a 30-case monitoring horizon. Robotic time and a supportive comparison of cases 1–5 versus 6–30 were secondary efficiency analyses. Postoperative radiographic measurements were performed by an observer blinded to the planned and robot-validated ROSA ® values. Agreement was quantified using Bland–Altman bias and 95% limits of agreement. Results Mean robotic time was 14.5 ± 2.4 min, mean non-robotic time was 8.9 ± 3.1 min (range, 4.1–15.6), and mean total surgical duration was 23.4 ± 2.3 min. The continuous LC-CUSUM score crossed the calibrated decision boundary (h = 4.5895) at case 6. Robotic time was 16.0 ± 3.8 min in cases 1–5 and 14.2 ± 2.0 min in cases 6–30 ( p = 0.787); corresponding total durations were 24.8 ± 3.8 and 23.1 ± 1.9 min ( p = 0.385). Bland–Altman bias for postoperative minus planned measurements was + 0.13° for tibial coronal alignment and + 0.10° for tibial slope; bias for postoperative minus robot-validated HKA was + 0.34°. All 30 postoperative radiographic measurements were within ± 2° of the planned target for both tibial coronal alignment and slope (100%; exact 95% CI, 88.43%–100%). Conclusions The predefined platform-specific temporal-performance standard was reached after six cases by an experienced high-volume UKA surgeon. Tibial planned-to-postoperative deviations were small in this consecutive series, while the findings remain conditional on the specified benchmark and require external validation.
No takes yet. Share an insight, caveat, or question.
Petrillo et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: