Key result
Retrograde transaortic retrieval successfully recovers embolized LAAC devices with safe same-session redo implantation.
Why the study?
Device embolization during LAAC into the left ventricle is rare but potentially life-threatening, and can interfere with the valvular apparatus or provoke ventricular arrhythmia.
Case Report (n=1)
Retrograde transaortic percutaneous retrieval of a left-ventricularly embolized LAAC device is feasible and can be followed by same-session redo closure in hemodynamically stable patients without structural injury.
Answers are grounded in this paper and cite their sources.
May permit percutaneous rescue and same-session redo LAAC in stable embolization cases; leaves open need for prospective safety data.
Background Left atrial appendage closure (LAAC) is an established stroke-prevention strategy for selected patients with atrial fibrillation who are unsuitable for long-term oral anticoagulation. Device embolization is rare but potentially life-threatening, particularly when the occluder migrates into the left ventricle, where it may interfere with the mitral or aortic valve apparatus or provoke ventricular arrhythmia. Case summary A 68-year-old woman with permanent atrial fibrillation, recurrent gross hematuria during anticoagulant therapy, chronic heart failure, and a high thromboembolic risk underwent LAAC under transesophageal echocardiographic guidance. Pre-procedural imaging showed a left atrial appendage ostium of 30 mm, a landing-zone diameter of 28 mm, and a depth of 32 mm. During the first deployment attempt with a 28/34-mm LAmbre occluder, the device showed insufficient anchoring and was not released. During attempted recapture and exchange, the partially deployed device embolized across the mitral valve into the left ventricular cavity, with a short self-terminating episode of ventricular tachycardia. Because the patient remained hemodynamically stable and real-time imaging showed no valvular entrapment or structural injury, retrograde transaortic percutaneous retrieval was performed using a 14-Fr steerable sheath and grasping forceps. After successful retrieval and immediate confirmation of valve integrity, a larger 32/36-mm LAmbre occluder was implanted in the same session with stable position and no residual peri-device leak. Discussion This case illustrates that, in carefully selected hemodynamically stable patients without valvular entrapment, left ventricular outflow tract obstruction, or structural injury, retrograde transaortic percutaneous retrieval of a left-ventricularly embolized LAAC device may be feasible. Same-session redo LAAC can be considered after successful retrieval when valve integrity is confirmed and the left atrial appendage anatomy remains suitable.
No takes yet. Share an insight, caveat, or question.
Tang et al. (2026) conducted a case report in Atrial fibrillation and left atrial appendage closure device embolization (n=1). Retrograde transaortic percutaneous retrieval and same-session redo LAAC was evaluated on Successful device retrieval and redo LAAC. Retrograde transaortic percutaneous retrieval of a left-ventricularly embolized LAAC device followed by same-session redo LAAC was successfully performed without structural injury or complications.