Answers are grounded in this paper and cite their sources.
Gel permeation chromatography (GPC) has been used to measure the molecular weight and characterize the number of arms per molecule for star-shaped styrene-butadiene block copolymers (SBS). Conventional GPC method based on polystyrene calibration standards measured a “linear-polystyrene-equivalent” molecular weight that was significantly different from the true molecular weight measured by a GPC hyphenated with a multiangle laser light scattering instrument. The comparison was made on four nearly monodispersed star-shaped copolymers, having a three-arm, four-arm, five-arm, or six-arm molecular structure. In all cases, the “linear-polystyrene-equivalent” molecular weights were higher than the true molecular weights for both the arm and the star molecule. The number of arms as determined based on the “linear-polystyrene-equivalent” molecular weights was consistently smaller than the true number of arms determined based on true molecular weights. The inaccuracy of the conventional GPC method was caused by the use of linear polystyrene calibration standards for characterizing a nonlinear copolymer molecule. The interpolation method developed previously for linear copolymers was unable to produce the accurate molecular weight for star SBS copolymers, and the deviation increases with an increase in the number of arms per molecule.
No takes yet. Share an insight, caveat, or question.
Chang et al. (2001) studied this question.