A photoacoustic 13 C 16 O 2 -laser-based gas analyser that permits low concentration measurements of pollutants in ambient air is described. The use of an isotopic 13 C 16 O 2 laser makes it possible to overcome the background absorption caused by atmospheric carbon dioxide. The resonant photoacoustic cell is placed inside the cavity of the laser. The minimum detectable absorption of 3 * 10 -9 cm -1 (concentration detection limit 0.3 ppbv for ammonia, 3 ppbv for arsine and 4 ppbv for phosphine) is reached at a time resolution of 30 s. The results of absorption measurements on individual lines of the 00 degrees 1-10 degrees 0 band of the 13 C 16 O 2 -laser for ammonia, arsine and phosphine in flowing nitrogen are reported. The concentration measurements for these substances in atmospheric air flows with different dilution ratios are carried out.
No takes yet. Share an insight, caveat, or question.
Starovoitov et al. (1993) studied this question.