Key result
Dual-gate Doppler mode measurements of internal carotid artery peak systolic velocity and carotid ratio were closely correlated with conventional duplex ultrasonography (r2=0.642 and r2=0.544; P<0.001).
Why the study?
This study was conducted to investigate the correlations of hemodynamic parameters measured to quantify carotid stenosis between standard duplex ultrasonography and the dual-gate Doppler mode.
Does dual-gate Doppler mode ultrasonography correlate with conventional duplex ultrasonography for hemodynamic measurements in patients with suspected or known carotid artery stenosis?
Observational (n=375)
No
Does dual-gate Doppler mode ultrasonography correlate with conventional duplex ultrasonography for hemodynamic measurements in patients with suspected or known carotid artery stenosis?
Dual-gate Doppler mode shows good correlation with conventional duplex ultrasonography for carotid hemodynamic measurements, offering a potential alternative despite minor precision discrepancies.
Dual-gate Doppler may serve as a feasible alternative for carotid velocity assessment; leaves open its accuracy for stenosis grading or intervention thresholds.
PURPOSE: This study investigated the correlations of hemodynamic parameters measured to quantify stenosis between the gold-standard duplex ultrasonography and the dual-gate Doppler mode. METHODS: Patients examined due to suspicion of carotid artery stenosis or for surveillance of known stenosis were invited to participate in this prospective single-center study. Upon acceptance, the hemodynamic characteristics of the carotid arteries were determined successively in standard duplex and dual-gate Doppler modes. The correlations between the two modes were analyzed by computing Pearson coefficients (r2) and Lin concordance coefficients (ρc). The degree of agreement between the two methods was visualized using Bland-Altman graphical representations. RESULTS: The correlation between internal carotid artery peak systolic velocity measured by standard duplex ultrasonography and dual-gate Doppler mode was good (r2=0.642). The same high level of correlation was observed for the carotid ratio (r2=0.544). However, the Bland-Altman graphical representation and the Lin concordance coefficients (ρc=0.75 and ρc=0.74 for the internal carotid artery peak systolic velocity and carotid ratio, respectively) showed that a lack of precision generated some discrepancies between the two measurement methods. CONCLUSION: Although some discrepancies were observed, the hemodynamic measurements were closely correlated between the two ultrasonography modes. Therefore, the dual-gate Doppler mode may have obvious advantages over conventional ultrasonography, offering interesting development possibilities.
No takes yet. Share an insight, caveat, or question.
Hersant et al. (2021) conducted an observational in Carotid artery stenosis (n=375). Dual-gate Doppler mode vs. Conventional duplex ultrasonography was evaluated on Correlation of internal carotid artery peak systolic velocity (ICA PSV) and carotid ratio (CR) (r2=0.642 for ICA PSV; r2=0.544 for CR, p=<0.001). Dual-gate Doppler mode measurements of internal carotid artery peak systolic velocity and carotid ratio were closely correlated with conventional duplex ultrasonography (r2=0.642 and r2=0.544; P<0.001).