Key result
Four automated natriuretic peptide assays demonstrated acceptable analytic performance, with overall analytic concordance values ranging from 84.3% to 95.9% compared to the Triage meter BNP assay.
The Access 2, ADVIA Centaur, AxSYM, and E170 automated natriuretic peptide assays demonstrate acceptable analytic performance and concordance with the Triage meter BNP assay.
May support automated assay adoption in labs; leaves open clinical outcome validation across populations.
Measurement of circulating B-type natriuretic peptide (BNP) and N-terminal proBNP (NT-proBNP) can identify patients with heart failure and guide therapy. The limit of detection, linearity, imprecision, method comparison, analytic concordance, and reference intervals of the Access 2 BNP (Biosite, San Diego, CA), ADVIA Centaur BNP (Bayer Diagnostics, Tarrytown, NY), AxSYM BNP (Abbott Diagnostics, Abbott Park, IL), and E170 NT-proBNP (Roche Diagnostics, Indianapolis, IN) methods were evaluated. The Triage meter BNP assay (Biosite) was the comparison method. Imprecision testing showed total coefficients of variation of 4.1%, 4.4%, 5.5%, and 0.8% for the Access 2, ADVIA Centaur, AxSYM, and E170, respectively. Relative to the Triage meter, method comparison revealed a slope of 0.96 and r = 0.95, a slope of 0.77 and r = 0.92, a slope of 1.13 and r = 0.94, and a slope of 8.8 and r = 0.80 for the Access 2, ADVIA Centaur, AxSYM, and E170, respectively. Overall analytic concordance values with the Triage meter were 95.9%, 92.9%, 92.4%, and 84.3% for the Access 2, ADVIA Centaur, AxSYM, and E170, respectively. All automated natriuretic peptide methods showed acceptable analytic performance.
No takes yet. Share an insight, caveat, or question.
Rawlins et al. (2005) studied this question. Automated natriuretic peptide assays (Access 2, ADVIA Centaur, AxSYM, E170) vs. Triage meter BNP assay was evaluated on Analytic concordance with the Triage meter. Four automated natriuretic peptide assays demonstrated acceptable analytic performance, with overall analytic concordance values ranging from 84.3% to 95.9% compared to the Triage meter BNP assay.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: