Key result
Central systolic blood pressure estimated using brachial oscillometry was similar to radial tonometry, with a mean difference of 0.36 mm Hg and strong correlation.
Why the study?
Does central systolic blood pressure estimation using brachial oscillometry correlate with radial tonometry in a Japanese population?
Cross-Sectional (n=155)
Does central systolic blood pressure estimation using brachial oscillometry correlate with radial tonometry in a Japanese population?
Brachial oscillometry provides central systolic blood pressure estimates that are highly correlated and similar to those from radial tonometry in a Japanese population.
Answers are grounded in this paper and cite their sources.
Brachial oscillometry may offer a convenient alternative for central SBP estimation; leaves open need for outcome validation before wider adoption.
Background: New devices have recently been developed using various features of the waveform derived from a brachial cuff for noninvasive estimation of central systolic blood pressure (SBP). Central SBP estimated from brachial oscillometry has never been compared with that estimated from radial tonometry in a Japanese population. Subjects and Methods: We recruited 155 Japanese volunteers (mean age 58 ± 16 years, range 18–99 years; 66.5% women) and estimated their central SBP using brachial oscillometry (Mobil-O-Graph) or radial tonometry (SphygmoCor). Results: The mean (standard deviation) peripheral SBP and central SBP measured with brachial oscillometry was 128 ± 18 mm Hg and 118 ± 16 mm Hg, respectively, while the central SBP estimated using radial tonometry was 119 ± 18 mm Hg. The mean (standard deviation) difference in estimated central SBP between brachial oscillometry and radial tonometry was 0.36 ± 5.9 mm Hg, and the central SBPs estimated using these devices were strongly correlated (r = 0.946, intraclass correlation coefficient = 0.940, p < 0.001). Conclusion: Central SBP estimated using brachial oscillometry was similar to that estimated from radial tonometry in a Japanese population.
No takes yet. Share an insight, caveat, or question.
Hoshide et al. (2018) reported a cross-sectional. Brachial oscillometry vs. Radial tonometry was evaluated on Central systolic blood pressure (SBP) (MD 0.36, p=<0.001). Central systolic blood pressure estimated using brachial oscillometry was similar to radial tonometry, with a mean difference of 0.36 mm Hg and strong correlation.